Standard Battery Charger; Maintenance; Maintenance Intervals; Cleaning And Disinfecting - Dräger X-plore 8000 Notice D'utilisation

Masquer les pouces Voir aussi pour X-plore 8000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26
en
|

Maintenance

4.2.2 Standard battery charger

Fault
Cause
General error or
defect
Status LED is
flashing red.
5
Maintenance
5.1

Maintenance intervals

Work to do
Clean and disinfect the device

Visual inspection

Replacing the O-ring at plug-in or bayo-
net-type hose connector
1) for gas-tight packed devices, otherwise every 6 months
5.2

Cleaning and disinfecting

CAUTION
Health hazard
The undiluted agents are damaging to health if they come into
direct contact with the eyes or skin.
► Wear safety goggles and protective gloves when working
with these agents.
NOTICE
Potential damage to components
► Only use the prescribed processes and the cleaning and
disinfection agents specified for cleaning and disinfecting.
Other agents, methods, dosages and contact times may
damage the components.
For information on the appropriate cleaning and
disinfecting agents and how to use them, please
refer to document 9100081 on www.drae-
ger.com/IFU.

5.2.1 Cleaning and disinfecting the device

1. Dismantle the device:
a. Separate breathing hose, facepiece and blower unit
from each other.
b. Disconnect the carrying system from the blower unit.
c. If available, dismantle any accessories (e.g. hose and
device sleeves).
d. Dismantle the splash guard cover and filter (see
"Replacing the filter", page 23).
22
2. Clean the facepiece according to the appropriate
Remedy
3. Clean the breathing hose and carrying system:
Re-insert the
rechargeable bat-
tery in the battery
charger. If the error
occurs repeatedly
have Dräger Ser-
vice check the bat-
tery charger and
rechargeable bat-
tery.
4. Clean and disinfect blower unit and splash guard cover
In cases of strong contamination, the blower unit can be
rinsed under running water as follows.
1. Make sure the rechargeable battery remains inserted.
2. Close suction inlet and tube connection with plug
1)
X
1)
X
5.3
X
5.3.1 Visual inspection
Check all parts thoroughly and replace damaged parts if
necessary. In particular, check the filter sealing surface and
O-rings of the blower unit for damage (e.g. scratches) or
contamination.
5.3.2 Replacing or charging the rechargeable
Risk of explosion, fire or chemical hazard!
► Do not remove, insert or charge rechargeable batteries in
► Keep rechargeable batteries away from sources of heat.
► Do not short-circuit the rechargeable battery contacts.
► Only use recommended rechargeable batteries.
Removing the rechargeable battery:
1. Fold up carrying system if necessary.
2. Push battery lock button. Ensure that the rechargeable
3. Remove rechargeable battery.
Inserting the rechargeable battery:
1. Fold up carrying system if necessary.
2. Position the two lower connection points of the
3. Fold the rechargeable battery in until it snaps audibly into
Instructions for Use.
a. Prepare a cleaning solution containing water and a
cleaning agent.
b. Clean all parts with the cleaning solution using a soft
cloth.
c. Thoroughly rinse all parts under running water.
d. Prepare a disinfectant bath of water and a disinfecting
agent.
e. Place all parts to be disinfected into the disinfectant
bath.
f. Thoroughly rinse all parts under running water.
g. Dry all parts in the air or in a drying cabinet
(temperature: max. +60 °C). Do not expose to direct
sunlight.
using disinfectant cloths.
Water must not enter the battery compartment.
(available as accessories).

Maintenance work

battery
WARNING
potentially explosive or flammable environments.
battery does not fall down.
rechargeable battery at an angle in the battery
compartment sockets.
place.