Measuring distance
4 to 6 cm
Once the thermometer has
detected the correct measur-
ing distance of 4 to 6 cm, the
measurement starts automati-
cally. The fever/distance indica-
tor now lights up continuously in
blue. After a few seconds, the
measurement appears on the
display (Fig. 12). The end of the
measurement is signalled by a
long beep.
If the measurement is less than 37.5 °C (99,5 °F), if the fever/
distance indicator lights up green and if
display to indicate no fever, this indicates that the body tem-
perature is in the normal range.
If the measurement is between 37.5 °C and 37.9 °C, the fever/
distance indicator lights up yellow. This indicates that the body
temperature is in an increased range (Fig. 13).
Measuring distance
NOT 4 to 6 cm
If the measuring distance is not
4 to 6 cm, the fever/distance in-
dicator flashes blue (the device
also beeps if the acoustic signal is
4-6
switched on) and
the display (Fig. 11). If the correct
measuring distance is not found
after 20 seconds, the measure-
ment ends automatically.
appears on the
46
Fig. 11
appears on
Fig. 12
Fig. 13