6. Scope of Delivery of the System Knee Joint

Description
NEURO TRONIC system knee joint (without figure)
AGOMET® F330, 5g (fig. 9)
orthosis joint grease, 3g (fig. 10)
assembly/lamination dummy (fig. 11)

7. Load Capacity

The load capacity results from the releveant patient data and can be determined by using the Orthosis Config-
urator. Use the system components determined by the Orthosis Configurator when producing an orthosis and
mind the recommended production technique. You will find information on the production techniques in the
section "Orthosis Production" under "Online Tutorials" on our website www.fior-gentz.com.

8. Tools for Assembling the System Joint

Tools
T8 hexalobular screwdriver/bit
T15 hexalobular screwdriver/bit
T20 hexalobular screwdriver/bit
torque screwdriver 1-6Nm
stripping tool
combination pliers

9. Assembly Instructions

The system joint is delivered fully assembled. All functions are checked beforehand. You have to disassemble
the system joint for mounting it in the orthosis and for maintenance. To ensure an optimal functioning, follow
the assembly instructions below. Secure all screws with the torque specified in paragraph 9.6.
When mounting the system joint, mind the basic alignment as it is essential for the later function of the
orthosis. Note the information on this in the manual of the multi-purpose device.
30
fig. 9
Quantity
1
1
1
1
fig. 10
System Width
16mm
x
x
x
x
x
x
fig. 11
20mm
x
-
x
x
x
x