Menu de configuration
Density measurement mode
Navigation
Description
Sélection
Réglage usine
Information
supplémentaire
Manual profile level
Navigation
Description
Entrée
Réglage usine
Information
supplémentaire
Profile density offset distance
Navigation
Description
Entrée
Réglage usine
300
Sous-menu "Profile density"
Navigation
Setup → Advanced setup → Sensor config → Profile density
Setup → Advanced setup → Sensor config → Profile density → Density measurement
mode
In normal measure mode, measures at specified positions. In compensation mode
measures using next integer value of drum turns to improve accuracy.
• Normal measure mode
• Compensation mode
Normal measure mode
Accès en lecture
Accès en écriture
Setup → Advanced setup → Sensor config → Profile density → Manual profile level
Sets the level position in the tank where the manual profile density operation starts.
–999 999,9 ... 999 999,9 mm
1 000 mm
Accès en lecture
Accès en écriture
Setup → Advanced setup → Sensor config → Profile density → Profile density offset
distance
Profile density offset distance [mm] is the distance between start point and first
measurement point.
0 ... 999 999,9 mm
500 mm
Operator
Maintenance
Operator
Maintenance
Proservo NMS80