Menu de configuration
Upper density offset
Navigation
Description
Entrée
Réglage usine
Information
supplémentaire
Middle density offset
Navigation
Description
Entrée
Réglage usine
Information
supplémentaire
Lower density offset
Navigation
Description
Entrée
Réglage usine
Information
supplémentaire
298
Sous-menu "Spot density"
Navigation
Setup → Advanced setup → Sensor config → Spot density
Setup → Advanced setup → Sensor config → Spot density → Upper density offset
Defines an offset value which is added to the measured upper density value.
–999,99 ... 999,99 kg/m³
0 kg/m³
Accès en lecture
Accès en écriture
Setup → Advanced setup → Sensor config → Spot density → Middle density offset
Defines an Offset Value which is added to the measured Middle Density Value.
–999,99 ... 999,99 kg/m³
0 kg/m³
Accès en lecture
Accès en écriture
Setup → Advanced setup → Sensor config → Spot density → Lower density offset
Defines an offset value which is added to the measured lower density value.
–999,99 ... 999,99 kg/m³
0 kg/m³
Accès en lecture
Accès en écriture
Operator
Maintenance
Operator
Maintenance
Operator
Maintenance
Proservo NMS80
Endress+Hauser