Proservo NMS80
Error value
Navigation
Prérequis
Description
Sélection
Réglage usine
Information
supplémentaire
Endress+Hauser
A
1
2
3
4
5
6
7
8
86
Principe de l' é valuation des limites
A
Alarm mode (→ 282) = On
B
Alarm mode (→ 282) = Latching
1
HH alarm value (→ 285)
2
H alarm value (→ 285)
3
L alarm value (→ 286)
4
LL alarm value (→ 286)
5
HH alarm (→ 286)
6
H alarm (→ 287)
7
L alarm (→ 287)
8
LL alarm (→ 287)
9
"Clear alarm (→ 288)" = "Oui" ou mise hors/sous tension
10
Hysteresis (→ 289)
Setup → Advanced setup → Application → Alarm → Alarm → Error value
Alarm mode (→ 282) ≠ Off
Defines the alarm to be issued if the input value is invalid.
• No alarm
• HH+H alarm
• H alarm
• L alarm
• LL+L alarm
• All alarms
All alarms
Accès en lecture
Accès en écriture
10
Operator
Maintenance
Menu de configuration
B
10
1
2
3
4
5
6
7
8
9
A0029539