10-4 Checking the Air Filter
When symptoms such as reduced Power Level or frequent clogging of powder are
observed, release water by pressing the tip of the Air Filter at the bottom of the Control
Unit (Fig. 59).
If sufficient Power Level cannot be obtained as a result of prolonged use, the VCP
Element may have clogged. If this happens, contact your Authorized NSK Dealer.
10-5 Replacing the fuse
1) Turn OFF the Power Switch.
2) Unplug the AC power cord from the wall outlet.
3) Remove the Fuse Holder located at the back of the Control Unit while keeping the
nails at both ends pushed inward (Fig. 60).
4) Replace the fuse with a new one.
5) Push the Fuse Holder back into the place.
* Refer to "21 Spare Parts List" for details about replacement Fuses.
• Only use NSK recommended fuses. Failure may occur if other fuses are used.
• If a fuse blows but you cannot establish the reason, contact your Authorized NSK Dealer.
CAUTION
NOTICE • If the Fuse Box does not come out easily, use a hard-tipped implement to push the nails while pulling the Fuse Box out.
29
Fig. 59
Fuse
Fuse Box
Fig. 60