6. Assembly
• Insert the lamp mount, with the burner in-
stalled, into the lamp housing and tighten it
with the screws (24.8).
• Put the lid down again. Plug in the contact
plug as far as it goes and retighten the
screws.
• Place the lamp housing in the incident light
lamp housing receptacle (25.1) and fasten it
with the clamping screw on the side.
• Connect the lamp housing to the power supply
(26.1).
Fig. 25 Rear side of stand
1
Incident light lamp housing receptacle
2
Transmitted light lamp housing receptacle
3
12 V 100 W connection for transmitted light (symbol:
4
12 V 100 W connection for incident light (symbol:
5
Lamp power cable
1
2
30
Fig. 24 106 z lamp housing (on the side, open)
1
2
3
4
5, 6, 7 Adjusting screw for x-y reflector
8
9
)
)
Fig. 26
1
5
3 4
Cover raised
Collector
12 V 100 W lamp or
gas discharge lamp in mount
Reflector (mirror)
Fastening screw for lamp mount
Socket for contact plug
1
2
3
8
Rear side of the ebq 100 supply unit
Lamp connection
1
4
9
8
5
6
7