14.3.5
Connections in interconnections
To avoid cross currents between conductivity and temperature output programmable
NOTE
controls, recording instruments and the like, of which the negative inputs are inter-
connected, must be inserted at the final position in the current loop.
ATTENTION
14.4
Test certificate
Electrical test:
Leakage test:
Material test:
Resistance to compression:
Resistance to temperature:
Leakage test:
Functional test/calibration:
417101109_LMIT08.doc
Under no circumstances must 4...20 mA current interface cards be used, which deliver
a voltage to supply two-wire transmitters. (Hazard of distruction of the current outputs
of
LMIT
08)
in compliance with EN 61010-1/VDE 0411, part 1
for degree of protection IP 67 in compliance with DIN 40050-9
All Materials are suitable for food acc. to BGA
- Transmitter housing: 1.4301 (AISI 304)
- Window: PMMA
- Sealing for housing: Thermoplast K
- Intermediate flange: Noryl GFN 2
- Flange for measuring cell: 1.4308, cast (AISI 304 SS)
- Body of measuring cell: E-CTFE
- O-rings: EPDM
- Temperature sensor 1.4404 (AISI 316 L)
max. 16 bar at 20° C
up to max. 140° C/3,5 bar, briefly
- Cold water test: 10° C, 6 bar, duration 15 min.
- Hot water test: 90° C, 6 bar, duration 15 min.
- 4 passages
- Temperature endurance test: 45°C,duration 24 h
- Cold water test: 10° C, 6 bar, duration 15 min.
- Hot water test: 90° C, 6 bar, duration 15 min.
- Cold water test: 10° C, 6 bar, duration 15 min.
- Current output temperature
- Current output conductivity
- Calibration: 200 μS/cm, 2/20/200 mS/cm, 2 S/cm
- Proper assembly