4.
Installation instructions
If possible the probe should be mounted in vertical running pipe pieces (see Fig. 4.1) that
distorting effects on the measured values caused by airbubbles are excluded.
Fig. 4.1
Fig. 4.2
At very high ambient temperature caused by build-up of heat (> 50 °C) and extreme
vibrations of the pipe (> 150 Hz, 20 m/s2) we recommend to use the
NOTE
mounted version with separate transducer. If you use the compact version the pipe has to
be fixed infront and after the
Fig. 4.3
The transverse hole in the measuring calotte must always be aligned parallel to the axis of
NOTE
the pipe, i.e. in the direction of flow of the medium. Deviations can have a distorting effect
on the measured values (see Fig. 4.3).
417101109_LMIT08.doc
The flow has to be from below up (see Fig. 4.1).
It is recommended to put the
the electrical connections point down.
If the transmitter has to be installed in horizontal
running pipes, it has to be mounted from below
(see Fig. 4.2).
LMIT 08
(see Fig. 4.3).
- 53 -
LMIT 08
the way that
LMIT 08
in wall-
Rev. 3-01.10