Setting The Assembly Factor; Pharma Version; Automatic Tcα Determination - Ecolab LMIT 08 Notice Technique

Transmetteur pour le mesurage sans électrode de la conductivité et de la température
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

8.5.2

Setting the assembly factor

The assembly factor is a measurement unit that shows the influence the installation situation
exerts on the constant of the measuring cell.
The assembly factor enables the
situations. The table below shows the necessary assembly factors for tank installation and
flow fittings, which are supplied with the
Fig. 8.33
Fitting
Tank mounting
flow DN 40
flow DN 50
flow DN 65
flow DN 80
flow DN 100
flow DN 150

8.5.2.1 Pharma version

Within the "Pharma" version, a tank factor of 1.007 has to be adjusted for the measurement
ranges 200 µS/cm, 2 mS/cm and 20 mS/cm in combination with the DN 40-100 flow fittings.
Factory settings of the Pharma version are 1.007.
8.5.3
Automatic TCα determination
Fig. 8.34
In order to ensure exact determination of the TCα, it is necessary to heat the measured item
from the reference temperature to the working temperature or cool the measured item from
NOTE
the working temperature to the reference temperature. The temperature must change by at
least 5 °C in order to determine the TCα.
With the alternating display the automatic TCα determination can be started or stopped by using
the SEL key exactly at the desired temperatures.
The "running" message flashes while the temperature compensation is being determined. TCα
determination can be interrupted at any time using the SEL key. The maximum running time for
the TCα determination is 1 hour
Display after completing the necessary temperature interval:
Fig. 8.35
If a fault occurs:
Fig. 8.36
417101109_LMIT08.doc
LMIT 08
Assembly factor
1,000
1,020
The assembly factor is set ex-works to 1.020.
1,020
1,020
Setting range: 0.900 to 1.100
1,020
1,020
1,020
calibration to be adapted to various installation
LMIT
08.
If the temperature behaviour of the
measured item is unknown, the TCα of the
measured item can be determined as
descibed below.
- 73 -