7.2

Electrical connection

ATTENTION
Fig. 7.11
PE-connection
Cable connection
*internal = compact version
*external = wall mounted version
1) Lines twisted to one another (white/green)
2) Lines twisted to one another (black/red)
7.2.1

Auxiliary power connection

According to EN 60204, Section 8, the
addition to the supply voltage connections X1/1,2.
With respect to the wall-mounted transmitter, the potential equalization connection on the
attachment flange can also be used to connect the protective conductor.
7.2.2

Connection of current outputs for conductivity and temperature

In order to avoid current signal interference, a screened line (eg. LiYCY) is recommended. The
shielding wire should be connected at the programmable controller to the PE connection. In the
transmitter
LMIT 08
7.2.3

Transducer connection

Introduce measuring line through the right cable inlet (3) on the
Connect to the terminal 7...15, according to core wire colour codes.*
Tighten threaded cable union
* Connect potential equalization (see Fig. 7.12).
417101109_LMIT08.doc
To ensure compliance with protective enclosure IP 67 specifications, the scope of
delivery includes various sealing inserts for the PG 13.5 mm threaded unions for cables
from 4 to 10 mm diameter. Unused PG unions must be sealed.
X1
Core wire colour code
internal*
Lapp-cable
1
-
2
-
3
-
4
-
5
-
6
-
7
white
white
8
brown
brown
9
black
black
10
red
grey
11
white
pink
12
white
black
13
white
green
14
brown
yellow
15
black
black
the shielding must be insulated.
external*
Ecolab-cable
Auxiliary power 24V AC/DC
Current output 0(4)...20 mA
for conductivity
Current output 0(4)...20 mA
for temperature
1)
white
1)
green
Transmitter coil
-
yellow
brown
Pt 100
blue
2)
black
2)
red
Receiver coil
-
LMIT 08
has a protective conductor connection in
- 62 -
Cable inlet
1
2
Measuring cell
connetion 3
LMIT
08.
Rev. 3-01.10