Factory Settings; Calibration Functions; Functional Choice In Calibration Mode - Ecolab LMIT 08 Notice Technique

Transmetteur pour le mesurage sans électrode de la conductivité et de la température
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

8.4.5.1 Factory settings

Conductivity measuring range:
Reference temperature:
TCα:
Current output conductivity:
Current output temperature:
Device number:
CIP 1 (standard instrument):
CIP 1: (Pharma version)
CIP 2-4 (option):
PROFIBUS DP/DPV1 (option):
TCα is set ex works to 2,0 %/K
NOTE
Please notice that with the "default configuration" function TCα is set to 0 %/K.
8.5

Calibration functions

Fig. 8.31
8.5.1

Functional choice in calibration mode

Fig. 8.32
417101109_LMIT08.doc
0..200 mS/cm
20°C
2,0%K
4...20 mA, corresponds to 0...200 mS/cm
4...20 mA, corresponds to 0...100°C
- - -
0...200 mS/cm, Tkα: 2,0 %/K
4...20 mA, corresponds to 0...200 mS/cm
0...200 µS/cm, Tkα: 2,35 %/K
4...20 mA, corresponds to 0...200 µS/cm
0...200 mS/cm, Tkα 2,0 %/K
4...20 mA, corresponds to 0...200 mS/cm
Slave Adresse: 3, Tsdrmin-time: 128
- 72 -
Rev. 3-01.10