Together with liquids some solid particles may be sucked in,
therefore it is necessary to have fi lters in order to protect the
motor and recover wanted particles.
Aspi-Jet 6 γ is provided, like all other models, with a debris fi lter 9
(Fig. C) which can be checked from outside the cabinet; Aspi-Jet
7 γ is provided with a fi lter (10) on the canister cover.
Filters must be cleaned every day.
In order to remove the fi lter 9, fi rst turn the unit on and aspirate
only air for some seconds so that hoses and manifold dry out;
disconnect the power supply line, lift the terminals from their seat
and remove manifold 4 (Fig. C) pulling the fi lter outwards by its
handle. To check fi lter 10 lift bent pipe union 11 (Fig. D). For Aspi-
Jet 6 γ , every evening, once disconnected the power supply line,
loosen the two rubber bands and remove lid: take out canister,
empty and clean (Fig. E).
For Aspi-Jet 7-8-9 γ the canister must be removed once a week
and cleaned under running water.
Canisters, covers and probes of all models once a week
must be accurately cleaned with a sponge and Puli-Jet plus
diluted in hot water.
Every evening, after cleaning the fi lters, it is important to aspi-
rate Puli-Jet plus diluted in hot water; to prepare the Puli-Jet
plus solution follow the instructions on the bottle.
Disinfection and cleaning should be carried out by means of Pulse
Cleaner (Fig. F). Once rinsing is completed, Aspi-Jet 7-8-9 γ carry
out drainage automatically, while for Aspi-Jet 6 γ the canister is
to be emptied once again.
Puli Jet dissolves blood and mucus and performs an an ti mi -
cro bi al action; if used regularly, every day with hot water, Puli-Jet
plus removes old scalings and disagreable smells.
Do not use detergents, even with reduced foaming, as as pi rat ed
air volume and turbulence may cause foam build up and damage
the suction unit, make it stop and produce disagreable smells.
O-rings (tightening rings) and sliding closures of terminals (Fig.
G) must be lubricated with spray silicon every 15 working days.
It is advisable to replace all hose of the plant (Fig. H), expecially
outside the unit, and terminals periodically for sanitary and func-
tional reasons (fl exibility and lightness of hose, smoothness of
sliding closures).
Main cleaning and main te nance operations
• After every patient, replace and disinfect the external hose
• After every surgical operation or any long operation: rinse
the appliance by aspirating hot water.
• At noon, before lunch time, clean the system with sanitizing
Puli-Jet plus.
• After each working day: clean fi lters, clean plant with disin-
fecting Puli-Jet plus and hot water.
• Once every fortnight clean the canister, draining valve and
probes; lubricate OR and sliding clo sures of terminals with
spray silicon.