Ordinary Maintenance - Cattani Maxi-Smart Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

• Ordinary maintenance

• Ordinary maintenance must be carried out only by a pro-
perly trained member of surgery staff.
• We recommend you pay special attention to all the dan-
ger signs and use protective glasses, gloves and disposa-
ble coverall.
Every day *
• Check the display for alarm warnings, if danger signs are
shown contact the technician.
Every night
• At the end of the day suck up through the dental units a
Puli-Jet plus new disinfectant with anti-scale (A) use Pulse
Cleaner (B);
• Remove power before getting close to the suction unit.
Untrained staff must not get close to the machine.
Every day
• Place the antifoaming tablets in the fi lters of the dental
units.
Periodically
• Keep the fi lter of the aspirator clean.
Occasionally
• Ensure that no build-up (or simply a thick layer of dust) is
preventing proper ventilation in the suction unit;
• Keep the plant room clear of material that is not related
to the machines, in particular fl ammable material and make
sure that there is not a possibility that corrosive, fl ammable
or explosive mixtures are formed.
* Every day only during the initial period, then occasionally.
ALLARME:
TEMPERATURA ELEVATA
B
A
39