Routine Maintenance - Cattani Micro- Smart Manuel D'utilisation

Masquer les pouces Voir aussi pour Micro- Smart:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

• Routine maintenance

Routine maintenance must be entrusted to specially in-
structed surgery staff.
• We recommend that special attention be paid to all dan-
ger signals, and that protective goggles, gloves and dispo-
sable overalls for personal protection be used.
Daily
(especially at the end of the working day
and/or several times a day according to need)
• Check for any possible alarm on the display. In case of
alarms, contact the technician.
• At the end of every working day aspirate a solution of
Puli-Jet plus disinfectant using the Pulse Cleaner.
• Clean the aspiration fi lters on the dental unit, collect the
waste, especially amalgam, according to the regulations
in force and place the Disinfectant Antifoam Tablets in the
dental unit fi lters.
• Disconnect the machine from the mains before any
maintenance intervention.
• Clean the aspirator's fi lter.
Periodically, according to need
• Make sure that the aspirator ventilation is not obstructed.
• Keep the plant room free from anything not related to the
machines, especially from fl ammable material. Make sure
that there is no possibility for the formation of corrosive,
fl ammable and explosive mixtures.
ALLARME:
TEMPERATURA ELEVATA
6
27

Publicité

Table des Matières
loading

Table des Matières