Before connecting the aspirator to the piping of the centra-
lized system, ascertain that aspiration piping is clean, as
heavy debris can damage the appliance.
Connect the PVC light grey aspiration tube (3) (supplied
with the machine) to the 30 mm Ø tube-holder (2) ("aspira-
ted fl uid inlet"). The other end of the same tube should be
connected to the aspiration piping (2b) coming from the
surgeries.
The black heat resistant exhaust air pipe (4b), fi tted with
a metal spiral, must be connected to the 30 mm Ø tube-
holder (4) ("exhaust air outlet"). Connect the other end of
the pipe to the antibacterial fi lter (5), passing preferably
through a silencer (5a) supplied with the aspirator. The hot
air coming from the antibacterial fi lter must be conveyed
outside.
Connect the 18 mm Ø tube-holder (6, page 27) to the liquid
drain pipe. In the version fi tted with Hydrocyclone the aspi-
rator is draining liquids by gravity and therefore fl uids can-
not be pumped upwards at all, and the pipe must be at the
same level as, or lower than, the waste fl uid outlet. The 10
mm Ø tube-holder (8, page 26) ("emergency drain") must
be connected to the liquid drain pipe. In the version wi-
thout Hydrocyclone the 11 mm Ø tube-holder (6b) must be
connected to the liquid drain pipe. The piping connecting
the machine to the aspiration and draining system should
be fl exible to damper the small vibrations produced by the
aspirator.
The aspiration piping should be run in the
fl oor and at a point near the aspirator it
should rise about 30 cm to reach the tu-
be-holder (2) (draw. A and B, page 101).
If Micro-Smart is installed at a level lower
than the surgeries, the aspiration piping
must not enter the centrifugal separator
perpendicularly. Rather, place a few me-
tres of piping horizontally on the same
level of the machine and then connect
it to the centrifugal separator inlet which
is at a higher level (using a fl exible pipe)
(draw.B, page 101).
After the installation is completed, con-
nect the power cable to the mains power
supply according to the regulation EN
61008-1.