Summary of main cleaning and routine maintenance
operations to be entrusted to purposely trained
surgery staff
Extraordinary maintenance operations to be entrusted
to trained service engineers, provided with original spare parts
Detergents damage aspirators, hypochlorite destroys them.
O-rings (seals) and the slides for the closure of terminals (ill.
I) must be lubricated with Lubri-Jet every 15 working days.
It is important to replace all fl exible tubes once a year (ill.
L): in particular, the external hose and terminals (ill. L) must
be disinfected for each patient and replaced every three
months for hygienic and operational reasons.
Summary of main cleaning and routine
maintenance operations to be entrusted to
purposely trained surgery staff
• After every surgical operation or very long treatment: rin-
se the appliance by aspirating water, preferably warm and
Puli-Jet plus new. Replace and disinfect external tubes and
terminals.
• After every working day: clean the fi lters; clean the equi-
pment with warm water and Puli-Jet plus new. Insert the
disinfectant antifoam tablets in the fi lter.
• Every 15 days: clean the canister, the draining valve and
probes; lubricate O-rings and slides closing terminals with
Lubri-Jet.
Extraordinary maintenance operations to
be entrusted to trained service engineers,
provided with original spare parts
• Periodical checks: the service engineer shall check up all
relays, functioning of probes and draining pump, siphons
and waste outlets, all internal pipes, and plastic and rubber
compo-nents subject to ageing; hoses and seals should be
replaced once a year. The appliance must be tested periodi-
cally from the current safety regulations point of view. Short-
circuit probes: make sure that suction stops and that the
drainage pump starts operating.
• The manufacturer is at your disposal to supply spare parts,
literature, instructions and anything that might be useful.
Our concessionaires, agents, authorised retailers and servi-
ce engineers are provided with split-up drawings, electrical
diagrams, instructions and updating about maintenance and
service.