• Faults report and possibile solutions

• Maintenance and periodic checks reserved
for authorized technicians
5) In case the appliance is treated with products which are different from the ones recommended
by the importer, the correct working of the equipment might be compromised and faults could
occur.
6) Check that there are no liquid leaks. In case of working anomalies or alarms, contact the
technician.
• Faults report and possibile solutions (draw. 6 page 45)
The amalgam container fi lling levels are the only alarms directly linked to the Hydrocyclone.
Below is information useful to inform the reader about possible problems that may cause the
aspiration to stop and that may compromise the working of the "Separatore Centrifugo Micro-
Smart" (Centrifugal Separator Micro-Smart) for the Hydrocyclone.
1) If the display (15) is off it means that there is no power supply: check that the machine switch
and the master switch are on, in case they are on but the machine is still off, contact the tech-
nician.
2) The text " Report Cod. S09, S10 or S11 " indicates that the AC 100 amperometric control
system has detected an anomaly in the centrifugal separator current absorption.
Such an anomaly could have different causes: a mechanical blockage or a drain obstruction (8),
(draw. 3 page 42) of the Hydrocyclone or a temporary voltage decrease.
Disconnect the appliance from the mains for a few seconds to reset the alarms and to start the
appliance again. In case the alarm appears again, contact the technician.

• Maintenance and periodic checks reserved for authorized technicians

• Electrical shock risk : also 230 V
• General danger sign.
• Biological danger, risk of infections from epidemic diseases.
• Compulsory direction of fl ow or of rotation.
Yearly maintenance (draw. 8 page 47)
1) Clean the suction piping by aspirating at least 2 l. of a solution prepared with a detergent-
disinfectant for dental aspirators.
2) Remove the amalgam collection container (5) as per the instructions of the paragraph "Re-
placement of the amalgam collection container" and check the working of the level sensor (13)
inserting a metallic piece between the two photodiodes. The Hydrocyclone control panel or the
display will show a visual alarm to indicate that the amalgam level has reached 95% or 100%.
3) Remove the Hydrocyclone (20) upper part, clean all its elements, especially the fi ve holes (2),
check the conditions of the 'O' rings and lubricate them with Lubri-Jet before mounting them
again.