Install the Hydrocyclone below the separator in the vertical position as follows:
Mount the bracket (10) without disassembling the Hydrocyclone, as shown in the drawing. The
Hydrocyclone must be easily accessible from every side. Leave a free space of at least 10 cm
under the amalgam container (5) so it can be easily removed.
The fi lter (14) fi tted to the centrifugal separator has a particle size not larger than 1.2 mm, pro-
tecting the Hydrocyclone from larger particles which may damage it.
• Mounting (draw. 5 page 44)
• Electrical shock risk : also 230 V
• General danger sign.
• Biological danger, risk of infections from epidemic diseases.
• Compulsory direction of fl ow or of rotation.
Distances
The maximum distance does not depend on the Hydrocyclone but on the suction pump, we
therefore recommend that you refer to the instructions of the suction pump supplied by the ma-
nufacturer.
Positioning
The Hydrocyclone can be installed in the following layouts:
1) installation on the same level of the dental units
2) installation on the lower level
The Hydrocyclone must be easily accessible from all sides. Leave a free space on the top and on
the front of the unit to facilitate maintenance procedures.
Liquids drain due to fall, therefore we recommend to provide a fall towards the sewage to facilit -
ate the draining. It is possible to place a small trap, provided that it does not obstruct the draining.
16
can be lethal.
can be lethal.