7.2 Identifying incorrect operation

In the event of incorrect operation, refer to the tables below to quickly identify and repair the non-complex
parts of the medical device.
If the incorrect operation is not described in the tables below, please contact your supplier or the After-Sales
team at Ultradent Products Inc..
Do not use the medical device if it appears to be damaged or faulty. Isolate the medical device and make sure
that it cannot be used.

7.2.1 No operation

Symptoms: the indicator light on the medical device is off and the medical device is not working.

Possible causes

No electrical current
Internal fuse not working
Faulty connection between the power cord and the
electrical wall socket
The medical device also has internal fuses F1, F2 and F3 that cannot be accessed by the user.

7.2.2 No spray

Symptoms: There is no water spray at the tip.
Possible causes
Flow configuration button on minimum
Faulty water pipe connection
Low water pressure
Blocked filter
Tip or file blocked
Incorrect choice of tip
Inadequate amount of spray
10
• check that the spray works correctly and that there is no leakage.
A damaged or blocked cartridge must be replaced.
Contact your electrician
Return to Ultradent Products Inc. After-Sales team
Connect the power cord to the electrical wall socket
Adjust the flow configuration button
Check the water inlet
Check the water system pressure
Clean or change the filter
Unblock the tip or file
Check the tip
Adjust the spray
J58349 • V1 • (14) • 03/2014 • NB77US010B