7.2.3 The power is not as expected

Symptoms: the tip does not vibrate at the expected frequency, the treatment is not progressing as normal and
is taking longer or is at a standstill.
Possible causes
Worn or distorted tip
Incorrect use: incorrect approach angle or
inadequate pressure on the tooth
Presence of liquid or moisture between the
handpiece and cord
Wear of the O'ring handpiece seal

7.2.4 Ultrasounds not working

Symptoms: the tip does not vibrate, vibration cannot be heard.
Possible causes
Tip loose
Faulty connector contact
Handpiece cord wire(s) cut

7.2.5 Water leakage

Symptoms: Water is leaking from one of the following places:
• between the base of the handpiece and its cord
Possible causes
Wear of 1.15 mm x1 mm O'ring handpiece seal
8 Technical specifications for the medical
device

8.1 Identification

Manufacturer
Name of the medical device
Replace the tip
Refer to the configuration table available at
www.satelec.com/documents
Thoroughly dry the electrical contacts
Replace the O'ring handpiece seal
Tighten the tip using the wrench
Clean the cord contacts
Return to Ultradent Products Inc. After-Sales Department to
replace the cord
Replace the seal using F12304 kit