Removal from packaging,
installation, connections

3.1 Unpacking your medical device

When you receive your medical device, check for any damage that may have occurred during transportation.
If you have received this medical device by mistake, please contact the supplier to arrange for it to be
collected.
If you have any questions or requirements, contact your supplier.
The Ultrawave™ includes the following items:
• a Ultrawave™ unit with non-detachable pedal cord, a non-detachable cord and a handpiece support ;
• tips and wrenches depending on selected options;
• a Quick Start Ultrawave™ [J58350].

3.2 Positioning the medical device

Place the control unit in the position that is suitable for your activity.
The medical device must be placed on a secure and flat surface or a surface with a maximum slope of 5
degrees.
Adjust the position of your medical device to correspond to your angle of vision and the characteristics of your
workstation, e.g. lighting or distance between the user and the medical device.
Ensure that you can access your medical device quickly.
Do not install your medical device near or on another device.

3.3 Installing cords

Check that the cords do not hinder the movement or free circulation of anyone.
Never rotate the handpiece connector on its cord as this can damage your medical device.
Never wrap the handpiece cord around the medical device.
Make sure that it is not possible to wheel over or walk on the different cords.
The cord attached to its handpiece must be easily accessible. Make sure that the cord is slack during use.
Do not put the medical device cords in a cable cover or a cable tray.

3.4 Connecting the medical device to the water system

The information below only applies to a medical device that needs to be connected to the water system
to operate.
Ask an approved dental installation technician to connect your medical device to the water system.
The water supply system pressure may vary throughout the day. The water supply system pressure must be
adapted to the values recommended for your medical device. It is very important to make sure that the
maximum pressure permitted for the medical device is never reached or exceeded. If in doubt, you are strongly
advised to install or arrange for installation of a water pressure limiting system.
The water supply system must comply with the quality criteria compatible with the practice of dental
treatments.

3.5 Connecting the medical device to the electrical network

Set the medical device to OFF position O and check that the power voltage is compatible with that indicated
on the medical device or its power plug. Next, connect the cord to the wall socket in compliance with the
standards in force in the country of use.
A different voltage would cause damage to the medical device and could injure the patient and/or user. Any
variation in the electrical network voltage or electromagnetic field that is non-compliant with the limits in
force, could interfere with the medical device's operation.
The medical devices' ground must be connected to a supply network equipped with a protective ground.
Do not plug the medical device into an extension lead and do not put the power cord in a cable cover or
cable tray.