− If the case of irrigation coming from the dental practice water network, clean and rinse the cord, the handpiece
and the tip (or file) for 1 minute after each use.
− Soiled devices should be separated from non-contaminated devices to avoid contamination of personnel or
surroundings.
− Wipe blood and/or debris from the tips or files to prevent it from drying onto the surfaces.
− Cover the tip or the file with a soft lint-free wipe dampened with purified water to prevent blood and/or debris
from drying.
Containment and transportation
Soiled devices must be transported separately from non-contaminated devices to avoid contamination.
Preparation for pre-disinfection/cleaning
It is advisable to treat devices immediately or as soon as possible after use. Satelec devices must be treated within
two hours of use.
Before cleaning, unscrew the tip or the file at the front of the ultrasonic handpiece after each use.
Pre-disinfection/Cleaning – Automated method
Pre-disinfection/pre-cleaning method
Note: The manual pre-disinfection/pre-cleaning method has to be performed for the tips prior to automated cleaning.
Equipment: soft-bristled brush, soft and lint-free swab, lint-free cloth, syringe, pipette and/or water jet, ultrasonic
cleaner, washer/disinfector, enzymatic or alkaline cleaning solution.
Step duration (minimum)
1 minute
1 minute 30 seconds
1 minute
10 minutes
1 minute
Page 4/5 • J02009 ● V1 ● (13) ● 11/2013 ● RB01US010A
Rinse soiled device under cold running water. Use a soft-bristled brush or a
soft and lint-free clean cloth to remove most of the contamination and debris.
Use a syringe, a pipette or a water jet with an enzymatic or alkaline cleaning
solution to rinse the tip cannulation.
Manually wash device in a freshly prepared enzymatic or alkaline cleaning
solution. Use a soft-bristled brush to remove soil and debris. Wash device
under water to prevent aerosolization of contaminants.
Rinse device under cold running water. Use a syringe, a pipette or a water jet
to rinse the tip cannulation.
In an ultrasonic cleaner, immerse the device in a freshly prepared enzymatic
or alkaline cleaning solution.
Thoroughly rinse device with distilled or purified water. Use a syringe, a pipette
or a water jet to rinse the tip cannulation. Repeat rinsing.
Repeat the pre-cleaning procedure until no visible contamination
remains on device.