Step duration
(minimum)
1 minute
10 minutes
1 minute
2 minutes 30
seconds
1 minute
Perform a final rinse of device using distilled or purified water.
Dry device using a soft, lint-free cloth or clean compressed air.
Inspection
• Wrenches must be inspected to check that no contamination remains, that they are not corroded, dulled,
discoloured or damaged.
• Before packaging and sterilizing the cleaned products, check they are clean, undamaged and function properly.
• Damaged wrenches must be discarded. Wrenches must not be lubricated.
Packaging
Use suitable packaging for sterilization, such as a Sterile Barrier System in compliance with ISO 11607. Care should be
taken to protect wrenches from contact with other objects that may damage their surface or the Sterile Barrier System.
Sterilization
Unless otherwise specified, non-sterile products can be re-sterilized using validated steam sterilization methods
(ISO 17665 or national standards).
Satelec recommendations for packed wrenches are as follows:
Cycle type
Saturated steam -
forced air removal
(pre-vacuum)
Dry times generally range from 20 to 60 minutes due to the difference in packaging materials, steam quality, device
materials, total mass, sterilizer performance, and varying cool-down time.
The distributor and manufacturer accept no responsibility for sterilization procedures performed by the customer that are
not performed according to these Satelec recommendations.
Rinse soiled device under cold running water. Use a soft-bristled brush or a soft and
lint-free clean cloth to remove most of the contamination and debris.
In an ultrasonic cleaner, immerse the device in a freshly prepared enzymatic or
alkaline cleaning solution.
Rinse device under cold running water.
Manually wash device in a freshly prepared enzymatic or alkaline cleaning solution.
Use a soft-bristled brush to remove soil and debris. Wash the device under running
water to prevent aerosolization of contaminants.
Thoroughly rinse device with distilled or purified water.
Repeat the manual cleaning procedure until no visible contamination remains on device.
Sterilization exposure
time
4 minutes
Page 2/5 • J81009 • V1 • (13) • 11/2013 • RA14US010A
Cleaning instructions
Sterilization exposure
temperature
132°C
Dry time
Minimum 20 minutes