When the irrigation system has been cleaned, perform the following operations:
1. disconnect the handpiece and refer to the handpiece cleaning, disinfection and sterilization protocols by Ultradent
Products Inc. J12949;
2. clean and disinfect the medical device as indicated in the chapter Cleaning and disinfection of the
medical device page 9
3. follow the instructions for accessory cleaning, disinfection and sterilization protocols as outlined in the
documents [J81009] and [J02009].

5 Cleaning, disinfecting and sterilizing

The instructions relating to the accessory cleaning, disinfection and sterilization protocols provided by
Ultradent Products Inc. have been approved for each medical device and accessory. The applicable guides are listed
in chapter Associated documentation page 4
They can be downloaded at the following addresses www.satelec.com/documents or www.ultradent.com
In all cases, the local regulations in force relating to the accessory cleaning, disinfection and sterilization
protocols take precedence over the information provided by Ultradent Products Inc.

5.1 Cleaning and disinfection of the medical device

The medical device must be in OFF or O stop position during cleaning and disinfecting procedures.
Refer to the instructions detailed in the chapter Cleaning the irrigation system page 8-9
Avoid using cleaning and disinfection products that contain flammable agents.
Otherwise, ensure that the product has completely evaporated from or that there is nothing flammable left on the
medical device and its accessories before switching it on.
Do not use abrasive product to clean the medical device.
Never apply sprays directly to the medical device to clean it. Always spray the product onto a wipe,
then clean the medical device.
The medical device control unit, handpiece cord and control pedal must be cleaned and disinfected daily.
Use alcohol-impregnated wipes for disinfection.

5.2 Cleaning and disinfecting accessories

Refer to the accessory cleaning, disinfection and sterilization protocols listed in the chapter Associated
documentation page 4.
6 Monitoring and maintenance of the
medical device
Before and after each use, check that the device and its accessories are not faulty in any way. This is
necessary to detect any insulation fault or damage. If necessary, replace damaged parts.
Monitor the cleanliness of the air inlets on the control unit to prevent any heating.

7 Maintenance

The only preventive maintenance the medical device requires is:
• checking of accessories;
• everyday cleaning, disinfection and sterilization procedures;
• cleaning.

7.1 Identifying incorrect operation

In the event of incorrect operation, refer to the tables below to quickly identify and repair the
non-complex parts of the medical device.
If the incorrect operation is not described in the tables below, please contact your supplier or the After-Sales team at
Ultradent Products Inc..
Do not use the medical device if it appears to be damaged or faulty. Isolate the medical device and make sure that it
cannot be used.
J58109 • V1 • (14) • 03/2014 • NB79US010B
9