Ultradent Products ultrawave XS Manuel D'utilisation page 35

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Step duration
(minimum)
30 seconds
1 minute
30 seconds
8 minutes
1 minute
Automated cleaning method
Step
Pre-washing
Washing
Neutralization
Rinsing
Drying
Thermal disinfection
For automated cleaning: thermal disinfection at 90°C for a minimum of 5 minutes.
Inspection
• Wrenches must be inspected to check that no contamination remains, that they are not corroded, dulled,
discoloured or damaged.
• Before packaging and sterilizing the cleaned products, check they are clean, undamaged and function properly.
• Damaged wrenches must be discarded. Wrenches must not be lubricated.
Packaging
Use suitable packaging for sterilization, such as a Sterile Barrier System in compliance with ISO 11607. Care should be
taken to protect wrenches from contact with other objects that may damage their surface or the Sterile Barrier System.
Rinse soiled device under cold running water. Use a soft-bristled brush or a soft and
lint-free clean cloth to remove most of the contamination and debris.
Manually wash device in a freshly prepared enzymatic or alkaline cleaning
solution. Use a soft-bristled brush to remove soil and debris. Wash the device under
running water to prevent aerosolization of contaminants.
Rinse device under cold running water.
In an ultrasonic cleaner, immerse the device in a freshly prepared enzymatic or alkaline
cleaning solution.
Thoroughly rinse device with distilled or purified water.
Repeat the pre-cleaning procedure until no visible contamination remains on
device.
Duration (minimum)
2 minutes
2 minutes
5 minutes
2 minutes
40 minutes
Page 4/5 • J81009 • V1 • (13) • 11/2013 • RA14US010A
Cleaning instructions
Cleaning/decontamination instructions
Cold tap water
Warm tap water (> 40°C); use an alkaline or
enzymatic cleaning solution
Warm tap water with neutralizer, if necessary
Rinse with warm deionised or purified water
(> 40°C)
90°C

Publicité

Chapitres

Table des Matières
loading

Produits Connexes pour Ultradent Products ultrawave XS

Ce manuel est également adapté pour:

J58109

Table des Matières