Possible causes
Tip loose
Faulty connector contact
Handpiece cord wire(s) cut

7.1.5 Water leakage

Symptoms: Water is leaking from one of the following places:
• between the base of the handpiece and its cord.;
• under the tank;
• under the medical device.
Possible causes
Wear of 1.15 mm x1 mm O'ring
handpiece seal
Wear of the tank seal
Worn irrigation cassette

7.2 Corrective Maintenance

In the event of faulty operation, the following corrective maintenance actions may be performed by the user.

7.2.1 Replacing the fuses

The medical device is protected by two fuses in the power connector.
To replace the fuses, perform the following operations:
• stop the medical device (position O);
• disconnect the power cord from the electrical network;
• disconnect the mains cord power cord from the power connector;
• insert the tip of a flathead screwdriver into the notch on top of the fuse holder to release it;
• remove the used fuses;
• replace the used fuses with fuses of the same type and same rating;
• place the fuse holder in its recess by pushing it until you hear a click that confirms it is in the correct position;
Tighten the tip using the wrench
Clean the cord contacts
Return to Ultradent Products Inc. After-Sales Department to
replace the cord
Replace the seal using F12304 kit
Contact the After-Sales Deparment to replace the tank
Replace the irrigation cassette using kit F62202 as detailed in
the chapter Replacing the irrigation cassette page 12
J58109 • V1 • (14) • 03/2014 • NB79US010B
Solutions