4. Installation Instructions
Connecting the water inlet hoses:
Connect the hot and cold water inlet hoses to the hot and cold water supply.

Drain requirements:

Connect the drain hose to the drain outlet. The drain hose can be attached to existing drain
lines using a 3.5 cm / 1.5" or larger standpipe / P-trap combination. If the hose is connected
directly to the drain line, fittings and adapters should not reduce water flow. The drain hose
should be attached to the main drain at a point no more than 1 metre /3 ft. above the base of
the HYDRIM. A floor drain is acceptable (check local codes).

Electrical requirements:

Connect the electrical cord to the power supply. The HYDRIM must have a dedicated circuit
equipped with a 15A fuse. The outlet must be grounded.

Additional Information:

The HYDRIM unit is heavy. Use a forklift or dolly and exercise caution when moving it.
The HYDRIM should be serviced on site.
The HYDRIM is equipped with an air gap / anti-suction device to prevent backflow of dirty
water into the water supply. No other air gap device is necessary.
If you need to extend the water inlet and drain hoses, ensure that you use commercial grade
plumbing hose.
The maximum length of the drain hose is 3.3 m / 13 ft.
Installing the water softener salt:
Unscrew the salt container lid and pour 1 litre / 1 quart of water into
the water softener. Add 1 kg / 2.2 lbs of water softening salt in the
same manner. Screw the salt container cap on tightly.
Handling and installing the HIP™ cleaning solution
The HIP™ cleaning solution is used in dilute form as a detergent for the cleaning
of surgical instruments in the HYDRIM range of washers and washer disinfectors. It is a
colourless and almost odourless liquid which is completely soluble in water.
The pH range of the concentrated solution is between 9.3 to 9.8 and the solution is
therefore alkaline. It is therefore advised that certain precautionary measures are observed
when handling and during the loading of the solution into the machine, and when removing
and disposing of the empty box.