Specifications; Serial Number Plate - SciCan HYDRIM L110w G4 Manuel De L'utilisateur

Masquer les pouces Voir aussi pour HYDRIM L110w G4:
Table des Matières

Publicité

8. Specifications

Full Unit Height: 850 mm / 33.5"
Low Profile Unit Height: 820 mm/32.3"
Width: 598 mm / 23.54"
Depth: 600 mm / 23.75"
Depth with door open: 1200 mm / 47"
Weight: 80 kg / 176 lbs.
Maximum Running Noise: 78 dB(A)
Hot and cold water connections: 3/4"
Inlet water pressure: 1-10 bar
Incoming hot water temperature: 50-70°C / 122-158°F
Drain: 3/4"
Water softener: 1 kg / 2.2 lbs salt capacity
Filling System: 5.5L / 1.5 gallon safety maximum
Drying System: Heater 1 kW
Electrical Connection: 208-240 V 60 Hz 15 A
Equipment pollution degree: Pollution Degree 2
Equipment installation category: Installation category II
Maximum relative humidity: 80% for temp up to 31°C / 88°F
Operating temperature range: +5°C to 40°C / 41-104°F
Max. altitude: 2000m / 6561.7 ft.
Mains supply: + / -10% of nominal

9. Serial Number Plate

When ordering supplies, spare parts or requesting service, please ensure that
the information contained on the serial number plate is available (Model number,
serial number etc.).
The serial number plate is located at the bottom left on the rear panel of
the HYDRIM L110w G4 unit. A small label is located on the left hand side of
the chemical door.
50% for temp up to 40°C / 104°F