6. Care and Maintenance

6.1 Replacing the HIP™ Cleaning Solution
The HYDRIM L110w G4 uses a 3.8L / 1 gallon container of HIP™ Cleaning Solution
(Part number CS-HIPL). Use only the recommended cleaning solution and read the
MSDS before inserting into the unit.
HIP™ cleaning solution is used in dilute form as a detergent for the cleaning of surgical
instruments in the HYDRIM washer. It is a colourless and almost odourless liquid that is
completely soluble in water. Some precautions must be taken to safely handle this
chemical. For more information, see Chapter 4 Handling and Installing the HIP™

cleaning solution.

NOTE: To avoid spillage, replace the solution container only when the message appears
and be sure to wear gloves.
When a red X appears on the water softener/detergent icon. Press on the icon to
determine if the problem is with the water softener or cleaning solution level. If the
cleaning solution must be replaced, a red X will appear next to it.
To order additional HIP™ Cleaning Solution, contact your local dealer. When the message
"no detergent" appears on the screen, the cleaning solution container is empty and is in
need of changing. To avoid spillage, replace the solution container only when the message
appears and be sure to wear gloves.
NOTE: Each cleaning solution container will last approximately 30 cycles,
depending on which cycles are used.
To replace the cleaning solution container, follow these steps:
Turn the power switch OFF.
1.
Open the cleaning solution door.
2.
Disconnect the cleaning solution connector. Remove any dried chemical by
3.
immersing the connector in a cup of water.
Remove the empty cleaning solution container and discard or recycle.
4.
30