4. Installation Instructions

To open the wash chamber door, plug in the unit,
turn it on and press the unlock symbol.
Alternatively, remove the screw in the kickplate
and pull the ring. See section 5.28.
Tools and supplies required for installation:
• Slot screwdriver
• Channel locks
Specifications:
Voltage: 208 – 240 V
Frequency: 60 Hz
Rated Load: 2.5 kW
Circuit breaker: 15 A
per phase
Weight: 80 kg / 176 lbs
Max. Running Noise: 78 dBA
Utility Hook-ups
Hose /
Length /
Cord
Diameter
1.9 m / 6 ft.
Hot Inlet
3/4"
1.9 m / 6 ft.
Cold Inlet
3/4"
1.9 m / 6 ft.
R/O Inlet
3/4"
1.5 m / 5 ft.
Drain
3/4"
1.8 m
Electrical
AWG 14-3
Levelling the HYDRIM:
The unit is standing on three supports: rollers (wheels) at the
back and two legs at the front.
1. Remove the front kickplate. Push the HYDRIM
into place while lifting the strap at the front to allow the unit
to roll on the rollers.
2. Adjust the front legs as required until the
HYDRIM is level.
3. The rear two legs are used only if the floor is uneven or
cannot provide support to the rollers.
4. Tuck the strap under the machine before
replacing the kickplate, so as not to interfere
with the exhaust air vents.
Full Unit Height: 850 mm / 33.5"
Low Profile Unit Height:
820 mm/32.3"
Depth: 600 mm / 23.75"
Depth with door open:
1200 mm / 47"
Weight: 80 kg / 176 lbs
Max. Running Noise: 78 dBA
Max. Distance
Water
Distance from Pressure
inlet / drain
1.5 m / 5 ft.
1-10 bar /
14.5-145 psi
1.5 m / 5 ft.
1-10 bar /
14.5-145 psi
1.5 m / 5 ft.
1-10 bar /
14.5-145 psi
Installation:
Installation should be undertaken by a
SciCan approved technician. The use of
an unapproved installer may invalidate
the warranty.
If the HYDRIM is installed in a Sterilization
Center, the manufacturer of the Sterilization
Center should allow a 10mm / 1/2" space
at the top, back and both sides of the HYDRIM.
This will facilitate installation, and service
access to the HYDRIM.
Do not move the HYDRIM into place by
maneuvering the open wash chamber door.
This may cause to door to become misaligned
and leak.
During installation, all consumables should
have been added to the machine as appropriate.
It is important that this be done before starting
the machine.
Shut-off
valve
Rear of Unit
Yes
Yes
Yes
power
connector
cold water
inlet hose
RO water
inlet hose
4
For safety reasons, do not install
the HYDRIM without the top cover.
connector
fuse holder
drain hose
hot water
inlet hose
ethernet
connector
printer

Publicité