Opening The Door In Case Of Power Failure - SciCan HYDRIM L110w G4 Manuel De L'utilisateur

Masquer les pouces Voir aussi pour HYDRIM L110w G4:
Table des Matières

Publicité

5. Instructions for Use

5.28 Opening the Door in Case of Power Failure

If there is a power failure, remove the kickplate of the HYDRIM. Remove the screw
with a screwdriver and lift the kickplate up and toward you. Locate the ring and pull
it to open the door. Exercise caution! There may be fluid remaining in the unit and
the instruments may be hot. Instruments that have not completed the cycle should
not be used and should be reprocessed. When replacing the kickplate, ensure that
the grounding wire is attached.
Ring
Grounding wire
Screw
Kickplate