6. Care and Maintenance

6.6 Annual maintenance

If the message "Preventative Maintenance Required. Call for Service" appears, the annual
maintenance is required. To continue using the HYDRIM, press OK. Call for service as
soon as possible. The annual maintenance should be done after 365 days have elapsed
since the initial installation or since the last service call was performed.
Daily tests and checks
Annually
Bi-Annually
To ensure that the ongoing efficacy of the machine is assured, SciCan recommends that
the following periodic testing protocols are observed.
IMPORTANT NOTES:
This document outlines the minimum requirements to ensure that the equipment
performs to specification. In some regions, local/national guidance documents may
require additional testing. Please contact your dealer or local medical authorities
for details.
Installation, commissioning, annual servicing MUST be undertaken by SciCan
approved technicians. Failure to maintain this equipment may invalidate the results of
the following testing regime.
When
SciCan-approved
SciCan-approved
Who
door lock check
wash arm rotation check
door seal check
By the user
load carrier check
check and clean chamber filters
visual examination of each load for residual soil.
By a
Annual maintenance schedule
as outlined in the service manual
technician
By a
Re-validation if required by local regulations.
technician