VISIOMED easyScan VM-ZX1 Mode D'emploi page 76

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 5
Guidance and manufacturer's declaration - electromagnetic
immunity. The device is intended for use in the electromagnetic envi-
ronment specified below. The customer or the user should assure
that it is used in such an
Immunity
IEC 60601
test
Electrostatic
±6 kV contact
discharge
±8 kV air
(ESD)
lEC 61000-4-2
Electrical fast
±2kV for power supply
transient/
lines
burst
±1kV for input/output
IEC 61000-4-4
lines
Surge
±1kV
IEC 61000-4-5
differential mode
±2kV common mode
Voltage dips,
<5% UT (>95% dip in
short
UT) for 0.5 cycle
interruptions
UT (60% dip in UT) for
and voltage
5 cycles
variations on
70% UT (30% dip in
power supply
UT) for 25 cycles
input lines
<5% UT (>95% dip in
IEC 61000-4-11
UT) for 5 sec
Power
3 A/m
frequency
(50/60Hz)
magnetic
field
IEC 61000-4-8
Note : UT is the a.c mains voltage prior to application of the test level.
150
Visiomed - EasyScan VM-ZX1 - User manual - 062016
environment.11111111111111111111
Electromagnetic
Compliance
environment -
level
guidance
±6 kV
Floors should be wood,
contact
concrete or ceramic
±8 kV air
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30 %.
Not
Mains power quality
applicable
should be that of a
typical commercial or
hospital environment.
Not
Mains power quality
applicable
should be that of a
typical commercial or
hospital environment.
Not
Mains power quality
applicable
should be that of a
typical commercial or
hospital environment.
3 A/m
Power frequency ma-
gnetic fields should be
at levels characteristic
of a typical location in
a typical commercial or
hospital environment
Guidance and manufacturer's declaration - electromagnetic
immunity. The device is intended for use in the electromagnetic
environment specified below. The customer or the user should as-
sure that it is used in such an
EN
Immunity
60601
Compliance
test
test
level
level
3 Vrms
Conduc-
150
ted RF lEC
kHz
3 V
61000-4-6
to 80
MHz
3 V/m
Radiated
80
RF lEC
3 V/m
MHz
61000-4-3
to 2.5
GHz
Visiomed - EasyScan VM-ZX1 - User manual - 062016
environment.11111111111111111
Electromagnetic
environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d = 1,2
P
P
d = 1,2
80 MHz to 800 MHz
d = 2,3
P
800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmit-
ter manufacturer and d is the recom-
mended separation distance in meters (m).b
Fields strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, a should be less than the com-
pliance level in each frequency range.b
Interference may occur in the vicinity of
equipment marqued with the following
symbol:
151

Publicité

Table des Matières
loading

Table des Matières