Guidance and Manufacturer's Declaration—Electromagnetic Immunity
The FreeStyle Comfort is intended for use in the electromagnetic environment specified below. The customer or
the user of the FreeStyle Comfort should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electromagnetic
environment –
±8 kV contact
guidance
±15 kV air discharge
IEC 61000-4-2
±2 kV for power supply
lines
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for input/output
lines
±1 kV line(s) to line(s)
±2kV common mode
Surge
on AC lines
IEC 61000-4-5
±1kV differential on
AC lines
±2kV common mode
on outdoor I/O lines
0% U
45°, 90°, 135°, 180°,
225°, 270°, 315°)
Voltage dips, short
interruptions and
0% U
voltage variations
on power supply
70% U
input lines
for 25/30 cycles (0
IEC 61000-4-11
0% U
250/300 cycles (0
Power frequency
(50/60 Hz)
3 A / m
magnetic field
IEC 61000-4-8
NOTE U
is the A.C. mains voltage prior to application of the test level.
T
NOTE In some cases, the Freestyle Comfort may reset after a significant Electrostatic Discharge, and may
require a Full Power Cycle to resume normal operation. See Troubleshooting section for more details.
30 - ENG
PN MN232-C4 J | User Manual
Compliance level
±8 kV contact
±15 kV air discharge
±2 kV for power
supply lines
N/A
±1 kV line(s) to line(s)
for 0.5 cycle (0°,
0% U
T
45°, 90°, 135°, 180°,
225°, 270°, 315°)
for 1 cycle (0
)
0% U
0
T
(30% dip in U
)
70% U
T
T
)
for 25/30 cycles (0
0
for
0% U
T
)
250/300 cycles (0
0
3 A / m
Electromagnetic environment –
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical professional healthcare
facility and home healthcare
environments.
Mains power quality should be that
of a typical professional healthcare
facility and home healthcare
environments.
for 0.5 cycle (0°,
T
Mains power quality should be that
of a typical professional healthcare
facility and home healthcare
environments. If the user of the
for 1 cycle (0
)
0
T
FreeStyle Comfort requires continued
operation during power mains
(30% dip in U
)
interruptions, it is recommended that
T
T
)
0
the FreeStyle Comfort is powered from
an uninterruptible power supply or a
battery.
for
T
)
0
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a professional
healthcare facility and home
healthcare environments.