FreeStyle Comfort
5. Turn on the FreeStyle Comfort by pressing the
power button and select your prescribed flow
setting utilizing the increase (+) or decrease (-)
flow buttons.
A brief audible alarm will sound to indicate the
device has powered on.
Note: When the FreeStyle Comfort is initially turned
on the default flow setting is 2. If the setting is
changed, the unit will power on to the previous
flow setting.
6. Position the nasal cannula on your face per the
manufacturer's instructions provided with the
cannula and breathe normally. You should feel
a puff of oxygen delivered through the nasal
cannula each time you inhale.
7. To change the flow setting, press the increase (+)
or decrease (-) flow buttons to the desired flow
rate. It is normal to hear a difference in sound as
you change the flow settings.
12 - ENG
PN MN232-C4 J | User Manual
autoSAT Pulse Volume (mL)*Setting
Breath
1
2
Rate
15
10.5
21.0
20
10.5
21.0
25
10.5
21.0
30
10.5
21.0
35
10.5
21.0
40
10.5
21.0
*Tolerance is +/- 15%
non-autoSAT Pulse Volume (mL)* Setting
Breath
1
2
Rate
15
14.0
28.0
20
10.5
21.0
25
8.4
16.8
30
7.0
14.0
35
6.0
12.0
40
5.3
10.5
8. Turn the FreeStyle Comfort off by pressing the
power button.
Note: The display screen will enter standby mode
and dim after 20 seconds. Press any button to wake
the screen out of standby mode.
9. For Advanced Features see Menu Navigation
Instructions
SPECIAL PRODUCT FEATURES
autoSAT
®
CAIRE's proprietary autoSAT feature allows the
FreeStyle Comfort to deliver a consistent volume
of oxygen (bolus) per breath, up to a maximum
breath rate. As the user's breath rate increases
during activities like walking, the autoSAT feature
adjusts the device operation to maintain a set bolus
volume.
UltraSense
®
This sensitive pulse dose trigger detects the
pressure change during your breath, and releases
the pulse of oxygen when you inhale.
autoDOSE™
Should a "no breath alarm" occur, this feature will
pulse oxygen at 20 breaths per minute at the unit's
current flow setting until the alarm state is resolved.
3
4
5
31.5
42.0
52.5
31.5
42.0
52.5
31.5
42.0
42.0
31.5
35.0
35.0
30.0
30.0
30.0
26.3
26.3
26.3
3
4
5
42.0
56.0
70.0
31.5
42.0
52.5
25.2
33.6
42.0
21.0
28.0
35.0
18.0
24.0
30.0
15.8
21.0
26.3