Alarm Conditions
Priority # Alarm
Description
1
System
Malfunction
Battery High
Temperature
2
System
Malfunction
High
Temperature
3
System
Malfunction
High
Pressure
4
No Breath
Detected
5
Low Battery
Shut Down
6
Warm Up
Display
FreeStyle Comfort
Audible
More Details / Suggested Action
Signal
1 beep
Unit will alarm when battery tempera-
every
ture is approaching safety shutdown
20 sec
threshold.
If shutdown threshold is reached, unit
will shut down. If external power is
connected, alarm screen will display and
audible alarm will continue for 2 min.
Test with battery at room temperature. If
problem persists, contact provider.
1 beep
Unit will alarm when internal tempera-
every
ture is approaching shutdown threshold.
20 sec
If shutdown threshold is reached, unit
will shut down and alarm screen will
display and audible alarm will continue
for 2 min.
Ensure there are no blockages on any
vents of the unit. If problem persists,
contact provider.
1 beep
If shutdown threshold is reached, unit
every
will shut down and alarm screen will
20 sec
display and audible alarm will continue
for 2 min.
If problem persists, contact provider.
1 beep
No breath detected for 15 seconds.
every
Activated after first breath or warm up
30 sec
screen is off, whichever comes sooner.
Check cannula for kinks or leaks. Increase
the sensitivity setting. If problem persists,
contact provider.
1 beep
Alarms until battery runs out of power
every
and unit shuts down.
30 sec
Charge battery.
none
Unit will start by auto-dosing until a
breath is detected or Warm up ends,
whichever is sooner.
PN MN232-C4 J | User Manual
ENG - 19