4 Usage

4.1 Safety information

CAUTION!
Overloading the bath lift can lead to damage to the
device or prevent lifting.
– Do not overload the bath lift and observe the
specifications on the identification label and in section
9.1 Dimensions and weight, page 22 with regard to
its load capacity.
CAUTION!
Danger of pinching fingers
– Do not insert fingers into the guide between the seat
plate and the backrest.
– Do not reach under the seat plate or between the
lifting scissors.
IMPORTANT!
Moisture may penetrate into a damaged hand control,
which could lead to a short circuit and fire.
– Always perform a visual inspection of the product for
external damage before each use.
– If the housing of the hand control is damaged or shows
cracks, do not use the hand control.
– Recharge the battery completely before each use, 3.4
Charging the battery, page 12.
1.
Refer to operating conditions in section 9.4 Environmental
Parameters, page 23.
1574469-A

4.2 Controlling the bath lift

The bath lift can be moved in the following directions:
The bath lift is controlled via the hand control, either by the user
or an assistant.
With a fully charged battery, the ORCA and ORCA F can
perform 5 bathing operations with a load of 140 kg. The
ORCA XL can perform 4 bathing operations with a load
of 170 kg.
E
D
IMPORTANT!
Risk of damage
– The hand control may only be used when the backrest
is properly installed and in the upright position.
A
B
C
Usage