Commissioning
3.4 Sensor Installation
Figure 3-6
Note
Minimum Ltn 18mm (0.75 in)
Ltn Menu Cell
This view only menu cell shows the distance in inches or millimeters between the front faces
of the sensors along the axis of the pipe. If you are mounting the sensors without a track or
spacer bar, you have to space them according to this value. Note that Ltn may be a negative
number for direct mount on very small pipes where the sensor spacing overlaps.
Installation Procedure (See figure for reference)
1. On a flat surface, attach the Spacer Bar to a Mounting Frame so that the Reference Hole
2. Slide the second Mounting Frame onto the other end of the Spacer Bar and align the
3. Wrap a Mounting Strap around the pipe. Make sure to position it so there is easy access
4. At the mounting location, place the Mounting Frame/Spacer Bar Assembly on the pipe so
26
Optional: On larger pipes, multiple lengths of
straps can be linked together to surround
pipe.
Mounting Strap positioned around Mounting
Frame
Sensor shown in the 9 o'clock position on pipe
Mounting Frame
Reflect Mount with Mounting Frames and Spacer Bar (Front View)
on the Spacer Bar fits over the metal post on the platform of the frame. Tighten the
clamping screw.
Number Index Hole with the metal post on the platform. Then tighten the clamping screw.
Ensure that the angled sides of both frames face away from each other.
to the Mounting Strap Adjusting Screw.
that it rests on the top of the pipe.
Space Bar Platform & Clamping Screw
Space Bar (Front View)
Metal Post
Mounting Strap Adjusting Screw
FUP1010 IP67 Portable Quick Start
Operating Instructions, 01/2013, CQO:QSG003 Revision 05