4. Press <ENTER>. Enter dimensions manually if pre-programmed dimensions do not
5. Press the <Left Arrow> and return to the main menu.
Select Liquid Class
1. Press the <Down Arrow> and scroll to [Application Data].
2. Press the <Right Arrow> to select [Liquid Class].
3. Press the <Right Arrow> again and scroll to desired liquid.
4. Press <ENTER> to save selection.
Select Pipe Configuration
1. Scroll down to [Pipe Config] and press the <Right Arrow>.
2. Select a configuration that approximates the conditions upstream of your Sensor
3. Press <ENTER> to save selection.
FUP1010 IP67 Portable Quick Start
Operating Instructions, 01/2013, CQO:QSG003 Revision 05
match application.
Note
The DN sizes listed in the [Select Pipe Size] menu option list are referenced to DIN Table
2448. After selecting pipe size, check pipe OD and wall thickness for correct dimensions.
Select from list.
mounting location. (Refer to the definitions below.)
Commissioning
3.3 Setting the Parameters
19