Siemens SITRANS FUP101 Mode D'emploi page 20

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Commissioning
3.3 Setting the Parameters
Note
To set English or Metric units: In the Meter Type menu, scroll to Meter Facilities Menu.
Press <Right Arrow> and select desired units. Press <ENTER> to select. Press <Left
Arrow> and <Up Arrow> to return to Meter Type menu.
Note
To select letters: Press <Right Arrow> to cursor and then press <Up/Down Arrow> to
select letters and numbers. Press <ENTER> when done.
4. Scroll to [Save/Rename Site]. Press <Right Arrow> then press <ENTER> to save site.
5. Press the <Left Arrow> and return to the main menu.
Select Pipe Class
1. Press the <Right Arrow> to select Pipe Class. Press <Right Arrow> again and scroll to
2. Press <ENTER> to select.
3. Pre-programmed Pipe Size and relevant pipe parameters will appear in menu cells. Press
18
Insert desired name (8 characters max.)
desired Pipe Class.
<Right Arrow> and scroll to desired pipe size.
FUP1010 IP67 Portable Quick Start
Operating Instructions, 01/2013, CQO:QSG003 Revision 05

Publicité

Table des Matières
loading

Table des Matières