WARNING
The procedures below must be carried out by suitably trained and qualified
personnel. Failure to do so may result in injury or an unsafe product.
Actions to be done by qualified personnel
Check that the bedstripper (linen shelf) (Optional) remains in its closed
position when maximum foot down tilt is applied
Check that the bed operates correctly using the backup battery as described
in the section "Battery test" on page 51.
Check operation the of castors, paying special attention to braking and
steering functions
Check that the calf section moves to the Fowler position when the thigh
section is raised
Check that the calf section latches securely in the horizontal (vascular)
position when manually raised.
Check that the bed extension locks securely in all three positions
Examine the power supply cord and mains plug; if damaged, replace the
complete assembly; do not use a rewireable plug
Examine all accessible flexible cables for damage and deterioration
Check all accessible nuts, bolts and other fasteners are present and correctly
tightened
Check any accessories fitted to the bed, paying particular attention to
fasteners and moving parts

Weighing system

Be sure to follow all local guidelines or regulations regarding weigh scale
verification and calibration.
The weigh system should be re-verified before the date of expiry shown on
the bed. Make sure the bed is re-verified every 12 months. If verification
does not pass, the weigh system should be re-calibrated.
Refer to the product service manual for verification and calibration
procedures or contact an approved Arjo service agent.
50
Fig. 26 - Weighing Calibration Expiry Label
1
Yearly