arjo Enterprise 9000X Notice D'utilisation page 41

Lits hospitaliers
Masquer les pouces Voir aussi pour Enterprise 9000X:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Initialisation
After connecting the bed to the electricity supply, the weighing system must
be initialised as follows:
1. Place the mattress, bed linen and all necessary accessories on
2. Press the Zero button once.
3. After a few seconds the display will read
4. The patient can now get onto the bed
5. The bed cannot be re-zeroed once a patient is on the bed. Use
WARNING
The weighing system is intended to provide reference data only. The weighing system
is not intended to provide readings upon which medication dosage decisions are
made.
CAUTION
The weighing system MUST be zeroed every time a new patient is to be placed on the
bed.
CAUTION
The weighing system MUST be zeroed every time a mattress is changed.
CAUTION
The weighing system cannot be zeroed if a mattress or accessories weighing more
than 35kg are fitted. To ensure the bed can be zeroed correctly, only use Arjo
approved mattresses.
the bed. The patient must not be on the bed at this point.
the Auto Compensation feature if adding or removing items from
the bed such as linen or accessories.
.
41