Decontamination

WARNING
Do not allow the mains plug or power supply cord to get wet.
CAUTION
Do not use abrasive compounds or pads, or phenol-based disinfectants.
Do not use jet stream cleaning or wash tunnels.
Do not remove grease from the actuator pistons.
NOTE
These instructions also apply to accessories but not to mattresses. For lifting straps and
handles, refer to the manufacturer's instructions supplied with the product.
The bed should be cleaned and disinfected weekly, and before a new
patient uses the bed.
Cleaning
Remove the mattress and all accessories from the bed.
The head/foot boards and mattress platform sheets should be removed
from the bed for cleaning.
Wearing suitable protective clothing, clean all surfaces with a disposable
cloth moistened in hand hot water and a neutral detergent.
Start by cleaning the upper sections of the bed and work along all horizontal
surfaces. Work methodically towards the lower sections of the bed and
clean the wheels last. Take extra care to clean areas that may trap dust or
dirt.
Wipe over with a new disposable cloth moistened with clean water, and dry
with disposable paper towels.
Allow the cleaned parts to dry before replacing the mattress.
Disinfecting
After cleaning the bed as described above, wipe all surfaces with sodium
dichloroisocyanurate (NaDCC) at a concentration of 1,000 parts per million
(0.1%) of available chlorine.
In the case of pooling body fluids, e.g. blood, the concentration of NaDCC
should be increased to 10,000 parts per million (1%) of available chlorine.
Use of other
Arjo recommends sodium dichloroisocyanurate (NaDCC) as a disinfectant
disinfectants
because it is effective, stable and has a fairly neutral pH. Many other
disinfectants are used in healthcare facilities, and it is not possible for Arjo
to test each one to determine whether it may affect the appearance or
performance of the bed.
If facility protocols require the use of a disinfectant other than NaDCC (e.g.
diluted bleach or hydrogen peroxide), it should be used with care and in
accordance with the manufacturer's instructions.