Error Codes
Error codes are shown on the display. They are used to indicate a problem
with the weighing system; this may be due to an operator error or a possible
fault condition.
The table below shows common error codes with their meanings and
possible causes.
Display
Zero on battery power
Auto Compensation
weight decrease more
than 100kg
Auto Compensation
weight increase more
than 100kg
A stable weight reading
was not obtained within
10 seconds
NOTE
If the display shows an error code other than those given above, refer to the product service
manual or contact an approved Arjo service agent.
Angle
When operating the controls for backrest angle or tilt, the weighing system
indication
display shows the approximate angle in degrees of the selected function.
Angles are shown relative to the floor, so the displayed values of backrest
angle and thigh angle change when the mattress platform is tilted.
The tilt angle is shown as a positive value for head down tilt and a negative
value for foot down tilt.
Cause
Connect bed to mains supply and re-initialise
the weighing system - see page 41
Replace weight removed from bed
Remove added weight from bed
Refer to the points listed in the section
"Accuracy" on page 40
Solution