arjo Enterprise 9000X Notice D'utilisation page 63

Lits hospitaliers
Masquer les pouces Voir aussi pour Enterprise 9000X:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601-1-2 test
Immunity test
level
Electrostatic
±2kV, ±4kV, ±8kV,
discharge (ESD)
±15kV air
EN 61000-4-2
±8kV contact
3V in 0,15 MHz to 80
Conducted
MHz
disturbances
6V in ISM and amateur
inducted by RF
radio bands between
fields
0,15 MHz and 80 MHz
EN 61000-4-6
80% AM at 1 kHz
Professional Healthcare
Radiated RF
environment
electromagnetic
3 V/m
field
80 MHz to 2,7 GHz
EN 61000-4-3
80% AM at 1 kHz
385 MHz - 27 V/m
450 MHz - 28 V/m
710, 745, 780 MHz -
Proximity fields
9V/m
from RF wireless
810, 870, 930 MHz - 28
communications
V/m
equipment
1720, 1845, 1970, 2450
EN 61000-4-3
MHz – 28 V/m
5240,5500, 5785 MHz -
9V/m
±1kV SIP/SOP ports
Electrical fast
±2kV AC port
transient/burst
100kHz repetition
EN 61000-4-4
frequency
Power frequency
30A/m
Magnetic field
50 Hz or 60 Hz
EN 61000-4-8
Compliance level
±2kV, ±4kV, ±8kV,
±15kV air
±8kV contact
3V in 0,15 MHz to 80
MHz
6V in ISM and amateur
radio bands between 0,15
MHz and 80 MHz
80% AM at 1 kHz
Professional Healthcare
environment
3 V/m
80 MHz to 2,7 GHz
80% AM at 1 kHz
385 MHz - 27 V/m
450 MHz - 28 V/m
710, 745, 780 MHz - 9V/
m
810, 870, 930 MHz - 28
V/m
1720, 1845, 1970, 2450
MHz - 28 V/m
5240,5500, 5785 MHz -
9V/m
±1kV SIP/SOP ports
±2kV AC port
100kHz repetition
frequency
30A/m
50 Hz
Electromagnetic environment
- guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material the relative
humidity level should be at least
30%
Portable and mobile RF
communications equipment should
be used no closer to any part of the
product, including cables, than
1.0m, if the transmitter's output
power rating exceeds 1W
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the compliance
level in each frequency range
Interference may occur in the
vicinity of equipment marked with
this symbol:
Mains power supply should be that
of a typical commercial or hospital
environment.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
a
.
b
.
63

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

E9x

Table des Matières