Preventive Maintenance Schedule; Daily Cleaning; Inspection / System Check-Out; Troubleshooting - arjo AtmosAir 4000 Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 62
Laundering Detachable Top Cover
1.
Unzip the top cover from the base for laundering.
Do not launder base as damage may occur.
2.
Recommended wash temperature for top cover is 60° C (140° F) for 15 minutes.
3.
Maximum wash temperature is 95° C (203° F) for 15 minutes.
4.
Tumble dry at 60° C (140° F) or air dry.
5.
Maximum drying temperature 80° C (176° F).

Preventive Maintenance Schedule

Preventive maintenance for the AtmosAir MRS consists of regular cleaning (see Care and
Cleaning) and an overall system check-out to be performed at the intervals described
below.
All components must be cleaned, disinfected and inspected after each patient's use
and before use by a new patient. Always use standard precautions, treating all used
equipment as contaminated. Institutions should follow local protocols for cleaning and
disinfection.

Daily Cleaning

The cover should be wiped daily with a mild soap and water solution. After wipe down, dry
surface with a towel.

Inspection / System Check-Out

Check each of the following before placing the AtmosAir MRS with a new patient:
1.
Check mattress surface for tears or cracking; do not use if tears or cracks are
present.
2.
Ensure mattress is free of stains and is not overly faded.

Troubleshooting

Do not attempt troubleshooting outside this guide or where the solution recommends to
contact Arjo. Any unauthorized service, modi cation, alteration or misuse may lead to
serious injury and / or product damage and will void all applicable warranties.
SYMPTOM
Mattress too firm upon arrival.
Mattress cover too wrinkled upon removal from
shipping container.
POSSIBLE CAUSE
Difference in altitude not sufficient to open valves.
Internal components have not accommodated
to environment. This does not affect inflation
or function.
SOLUTION
Apply weight to mattress to open valves.
Let mattress accommodate for 24 hours. If problem
continues, contact Arjo for assistance.