• If multiple Unity 2 Control Units or Docking Stations are in house, care must be taken not to mix
them. A particular Control Unit may only be used with the Docking Station it has been calibrated
with. If the Control Units and/or the Docking Stations are mixed up, the resulting system will be out
of calibration.
• Annual calibration of all transducers, reference microphones together with the Control Unit is re-
comended to maintain the performance of the system. Follow the guidelines of the local authorities
for proper calibration and servicing of audiological medical equipment.
• The transducers (headphones, bone conductor, measuring microphones etc.) supplied with the
instrument are calibrated to this instrument – exchange of transducers or the Control Unit requires
a new calibration.
• After calibration turn the Control Unit off and on again to activate the new calibration settings.
• An annual safety check must be carried out by qualified personnel authorised by Siemens.
• For a description of the main program functions refer to the Software Short Guide. A more detailed
description of the functionality is available in the Software Reference Manual. Both are available as a
PDF copy from the Unity Installation CD.

4. Installation of Unity 2 system

4.1 Description of the system

4.1.1 Content of the shipment

The Unity 2 system is shipped in two carton boxes which are particulary made to protect the devices during transporta-
tion. Please retain these boxes for any return shipment necessary.
Depending on the individual order, there are specific accessories packed together with the system. In case e.g. no
Docking Station was ordered, the shipment only comes in one box.
Please refere to the particular packing list included in the box to check the shipment is complete.

4.1.2 Control Unit

The Control Unit is the main part of the Unity 2 system
containing all the electronics and connections necessary
to run Audiometry, Real-Ear Measuremets and Hearing
Instrument Analyser. It includes the main power supply.
Inside the box of the Control Unit all transducers, micro-
phones and cable as well as the accessories for the Real
Ear Measurement and the Analyser are packed together
in a separate box.
The headphones and transducers included have been factory calibrated with the particular Control
Unit and are marked with the particular serial number of that Control Unit. Be sure to not mix the
accessories from other systems to maintain the calibrated status of the particular system.