RF Welded AtmosAir MRS (Detachable top cover)
1.
If a powered MRS, unplug pump before cleaning.
2.
Push bed linens to center of mattress to wipe down. Remove to launder top cover
or wipe down as described below.
The RF Welded AtmosAir MRS has a detachable top cover that can be
laundered. However the base is cleaned by wipe down methods only.
3.
Wipe and rinse any soiling from the mattress surface and base. Use 1000 ppm
chlorine or 70% alcohol.
4.
Rinse with a wet towel to remove chemicals.
5.
After wipe down, dry surface with towel.
6.
7.
Clean pump and tubes (if needed) by wiping with a damp cloth.
Laundering Detachable Top Cover
1.
may occur.
2.
Recommended wash temperature for top cover is 60° C (140° F) for 15 minutes.
3.
Maximum wash temperature is 95° C (203° F) for 15 minutes.
4.
Tumble dry at 60° C (140° F) or air dry.
5.
Maximum drying temperature 80° C (176° F).

Cleaning AtmosAir Pump and Tubes

1.
Ensure pump is unplugged.
2.
Wipe surface of pump and tubes with coarse cloth using an approved disinfectant
pump with cleaning solution.
from pump wearing rubber gloves or while unit is unplugged to avoid
components in area of spill. Fluids remaining on controls can cause
corrosion, which may cause components to fail or operate erratically,
3.
4.
Disinfect pump and tubes with a chlorine solution (mixed according to
the instructions in the Care and Cleaning section). Using a clean cloth,
wring out excess solution until cloth is damp. Wipe pump surface and tubes
with damp cloth.
5.
Allow to air dry.
10
Do not ood any part of the

Publicité